Intended use
To control the photometer function of the 4 measuring channels and the 4 mixer motors of APACT 4004.
Contents
3 × 2 ml vials with different solutions, ready to use.
| Photometer control | Solution |
|---|---|
| 1 | PPP |
| 2 | PRP |
| 3 | 50 % Solution |
Procedure
APACT 4004: Select assay QC-Test on the APACT and LPC-Software.
Cuvettes: Shake the Photometer-Controls very well before use. Pipette 200 µl of each Photometer-Control into a cuvette. Use the cuvette of Photometer-Control 1 for all 4 channels to store PPP-value. Use the cuvette of Photometer-Control 2 for all 4 channels to store PRP-value. Use the cuvette of Photometer-Control 3 for all 4 channels to measure the 50 % value.
Expected values
40 – 60 % max. Aggregation. The differences from the mean value of channel 1 -4 should not be higher than +/- 4 % max. Aggregation.
Example: Max. Aggregation %
| Channel 1 | Channel 2 | Channel 3 | Channel 4 |
|---|---|---|---|
| 47,0 | 49,0 | 47,0 | 49,0 |
Mean % Aggreg. = 48,0 (= range 44,0 – 52,0). QC = okay, all values are in the range.