Service Photometer Control Kit

Automatically transferred from the PDF. Please compare details with the original document.

Intended use

To control the photometer function of the 4 measuring channels and the 4 mixer motors of APACT 4004.

Contents

3 × 2 ml vials with different solutions, ready to use.

Photometer controlSolution
1PPP
2PRP
350 % Solution

Procedure

APACT 4004: Select assay QC-Test on the APACT and LPC-Software.

Cuvettes: Shake the Photometer-Controls very well before use. Pipette 200 µl of each Photometer-Control into a cuvette. Use the cuvette of Photometer-Control 1 for all 4 channels to store PPP-value. Use the cuvette of Photometer-Control 2 for all 4 channels to store PRP-value. Use the cuvette of Photometer-Control 3 for all 4 channels to measure the 50 % value.

Expected values

40 – 60 % max. Aggregation. The differences from the mean value of channel 1 -4 should not be higher than +/- 4 % max. Aggregation.

Example: Max. Aggregation %

Channel 1Channel 2Channel 3Channel 4
47,049,047,049,0

Mean % Aggreg. = 48,0 (= range 44,0 – 52,0). QC = okay, all values are in the range.