What are you looking for?

Find tests, products, news and congresses across our website.

Your selection

Insert request

Complete request

Product information

Emicizumab Calibrator

Automatically transferred from the PDF. Please compare details with the original document.

Emicizumab calibrator is a calibrator plasma for the calibration of a modified one-step FVIII activity assay for the determination of active emicizumab concentrations in citrated plasma from a clotting time measurement.

Test principle

Emicizumab calibrator enables the quantification of the degree of correction of the APTT-based one-step assay by emicizumab in the sample. The calibrator is diluted in different concentrations. Each concentration is combined and incubated with FVIII-deficient plasma and APTT reagents. The reaction is started by the addition of CaCl2. The clotting time is then measured. The correction level of APTT is related to the emicizumab activity and is converted to µg/ml using the calibration curve.

Reagent and stability (reconstituted)

The reagent must be stored unopened at +2-8°C and can be used until the date indicated on the label.

ReagentQuantityStability / note
Emicizumab Calibrator5 x 1 ml8 h at 2-8 °C 4 h at 23-25 °C
Article number
151-201-CE

Date: 30.06.2014

© Haemochrom Diagnostica GmbH

Open original PDF

HAEMOCHROM DIAGNOSTICA

PIONEERING DIAGNOSTICS

We are here for you

Do you have a question about our products or services? We look forward to hearing from you.

Write to us

Send us your question or request using the form.

Fields marked * are required.
Loading…
Loading the web debug toolbar…
Attempt #