Product information

ELISA t-PA Actibind®

Automatically transferred from the PDF. Please compare details with the original document.

Complete ELISA-kit for quantitative determination of the activity of human tissue-type plasminogen activator (t-PA) as well as the total human t-PA concentration in human plasma.

Test principle

The diluted samples are incubated in micro wells coated with monoclonal anti-t-PA-antibodies. After binding of t-PA to the capture antibody, unbound substances are removed by washing and a mix of Glu-Plasminogen and plasmin-substrate is added. After a photometric activity determination the plate is washed leaving the bound t-PA on the plate. A secondary, peroxidase conjugated, anti-t-PA-antibody is added which recognizes bound t-PA. After a washing step, added substrate reacts with the conjugate. The intensity of the resulting colour is proportional to the content of t-PA in the sample.

Reagents and stability (after opening)

ReagentQuantityStability / note
Coated micro wells12x8expiry date at 2-8°C
Wash buffer (10-fold)1 vial6 months at 2-8°C
Incubation buffer1 vial2 months at 2-8°C
Standard1 vial6 months at -20°C
Conjugate1 vial6 months at 2-8°C
TMB-substrate1 vialexpiry date at 2-8°C
Plg.-act.-det. mix1 vial2 hours at 2-8°C 2 weeks at –70°C
Plg.-act.-det. mix dil.1 vialexpiry date at 2-8°C
Number of determinationsArticle number
96TC16000

Date: 15.11.11

© Haemochrom Diagnostica GmbH

Open original PDF

HAEMOCHROM DIAGNOSTICA

PIONEERING DIAGNOSTICS

We are here for you

Do you have a question about our products or services? We look forward to hearing from you.

Write to us

Send us your question or request using the form.

Fields marked * are required.
Loading…
Loading the web debug toolbar…
Attempt #