Fibrinogen Reagent

Automatically transferred from the PDF. Please compare details with the original document.

Reagent for determination of fibrinogen according to Clauss with very high thrombin concentration. Thereby the reagent becomes insensitive to heparin.

Test principle

Diluted citrate plasma is mixed with Thrombin (~ 80 I.E./ml) in excess and the coagulation time determined. The fibrinogen content is plotted against a calibration with normal plasma to find the result of the sample.

Reagents and stability (after opening)

Fibrinogen Reagent Kit:

ReagentQuantityStability / note
Fibrinogen Reagent5 x 2 ml8 hours at RT 1 Day at 2-8°C 1 Month at -20°C
Coagulation Reference1 x 1 ml4 hours at RT 8 hours at 2-8°C 1 Month at -20°C

Analyzer applications: on request

VariantPack sizeArticle number
Fibrinogen Reagent Kit—5138005
Fibrinogen Reagent5 x 2 ml5138085
Fibrinogen Reagent5 x 5 ml5138080